Long-term stability · Human factors
Implanted brain-computer interface functionality during nighttime in late-stage amyotrophic lateral sclerosis
- Authors
- Leinders S … Vansteensel MJ13 authors
- Institution
- UMC Utrecht
- Publication
- Scientific ReportsSci Rep 2026;16(1):14001 · 18 Mar 2026
- Status
- Primary source
The 60-second view
- What problem?
- A communication BCI for someone who is locked in has to work around the clock, not least to call for help at night. Nobody had tested one during sleep.
- What did they do?
- Analysed the night-time signals of one long-term user of an implanted ECoG BCI, ran the daytime decoder on them, then built a night mode and followed it in use.
- What changed?
- The signal itself is different at night: power in the low and high frequency bands was higher and more variable than by day, enough to set off a daytime decoder hundreds of times an hour.
- Why does it matter?
- A caregiver call that goes off 13 times an hour by mistake is one a household learns to ignore. For an implant meant for use at home, how it behaves during sleep belongs in the evidence.
Method
- Participants
- One person with late-stage ALS, a long-term user of an implanted ECoG BCI
- Interface
- Fully implanted electrocorticography system
- Signal
- Power in low and high frequency bands
- Task
- Click-based control and a call to the caregiver, by day and at night
- Decoder
- The daytime click decoder, and a separate night mode
Results
- 100%
- Nights with unintended activations
- Daytime decoder applied to night data
- 245 per hour
- Unintended clicks at night
- Daytime decoder
- 13 per hour
- Unintended caregiver calls at night
- Daytime decoder
- 79%
- Nights the night mode ran without error
- Over about 1.5 years
Strength of evidencePeer reviewed, one participant with late-stage ALS, with the night mode followed for about a year and a half.
Figures are quoted from the published abstract and were checked on 11 Oct 2026.
Limitations
- A single participant, with one implant and one decoder design.
- The night mode was free of errors on about four nights in five, not on all of them.
What this could enable
Treating round-the-clock reliability as a requirement with tests of its own, separate from daytime accuracy.
Primary source
- 01
Leinders S et al. Sci Rep 2026;16(1):14001 (opens in a new tab)
Peer reviewedScientific Reports18 Mar 2026Checked against source 11 Oct 2026
DOI 10.1038/s41598-026-44228-7